(Clinical Chemistry. 2006;52:553-554.)
Why Commutability Matters
1 Department of Pathology Virginia Commonwealth University
Richmond, VA
2 Division of Laboratory Sciences National Center for
Environmental Health Centers for Disease Control and Prevention
Atlanta, GA
3 Wadsworth Center for Laboratories and Research New York
State Department of Health and School of Public Health State University
of New York at Albany Albany, NY
aAddress correspondence to this author at: PO Box 980286, Richmond, VA 23298-0286. Fax 804-828-0375; e-mail gmiller@....
| The first 20% of the full text of this article appears below. |
A fundamental goal of laboratory medicine is that results for patients’ samples will be comparable independent of the medical laboratory that produced the results. Routine measurement procedures of acceptable analytical specificity that have calibration traceable to the same higher-order reference material or reference measurement procedure should produce numerical values for clinical samples that are comparable irrespective of time, place, or laboratory generating the result.
The term "commutability" was first used to describe the ability of a reference or control material to have interassay properties comparable to the properties demonstrated by authentic clinical samples when measured by more than one analytical method (1)(2)(3). More recent metrologic documents expand the concept; they describe commutability as the equivalence of the mathematical relationships between the results of different measurement procedures for a reference material and for representative samples from healthy and diseased individuals (4)(5).
A reference material is typically used to establish or verify the
traceability of a measurement procedure to a value assigned to
the material that represents the best estimate of the correct, or
true, value. Concentration or activity values are assigned to
reference materials by higher-order measurement procedures, when
possible, or alternatively by a consensus process when reference
measurement procedures are not available. When a reference material
is intended to be measured by a routine clinical method, commutability
must be validated